Make: Zenotrex Medico
Model: ZPL-A22R

VERTICAL AUTOCLAVE FULLY AUTOMATIC

Vertical autoclaves are also known as top loading sterilizers that come in cylindrical shape. The working of vertical autoclave is simple to understand; to open, a radial locking mechanism is fitted at top of the lid; a foot lever (pedal) is attached at the bottom. User has to press the pedal and lift the lid. Vertically designed chamber provides guaranty of safe and efficient sterilization even in confined spaces.

FEATURES:

  • Capacity:- 20-150 Liters
  • Equipped with joint less silicon rubber gaskets
  • Simple and Easy to load and unload
  • Completely Digital monitorable, microprocessor-controlled One-touch operation, constant management and monitoring of autoclave functions and parameters
  • Steam collection bottles to remove most of the steam during operation.
  • Ware inlet and outlet valve with inbuilt 
  • Drain valve for cleaning or changing with fresh water
  • Constructed with appropriate stainless steel with superior corrosion resistance to water and steam
  • High temperature and pressure resistant silicon gasket
  •       Start-timer setting range: 1-99 hours in each operation mode
  •       Floor mounted & self stand.
  • Preset timer: atleast upto 9 hours.
  • A water level gauge. Analog gauges for measuring inner and outer steam pressure.
  •      PT100 type temperature sensor.
    SPECIFICATIONS:

 Working Pressure & gauge

 Variable 15- 20 PSI(0-127kPa) with Analog  Pressure Gauge(Maximum  Pressure: 147 kPa) (Optional 4bar)

Working Temperature

100°C to 140°C(Facility of warming temp range 40°C to 60°C and Dissolving Temperature range 40°C to 100°C.)

Internal Chamber

304/316 SS

External Chamber

Mild Steel (304 stainless Steel – Optional)

Door

Foot lifted with safety and interlock device stainless steel sheet of 304 grade (min.5 mm thickness) Radial locking type. Elastomeric rubber gasket suitable to withstand temperature upto 140°C and pressure upto 2 bar.

Insulation

Glass wool insulation

Heating device (steam generator)

inside the jacket made up of heavy gauge stainless steel, separated from the chamber with upto 60°C to 150°C drying temperature.

Sterilization Period

25 to 30 Minutes

Time display Range

a. Sterilizing Mode: 1-240 minutes or continuous

b. Sterilizing-warming mode: 1-240 minutes (sterilizing)/48 hours (warming)

c. Heating mode: 1-8 hours

d. dissolving time upto 48 hours

LED display

LED display for temperature and remaining time

Power Supply

220-240 Volts, 50/60 Hz, single phase

      Additional Feature

Leakage breaker:

a. Rated breaking current: 20A (120V), 15A (220/230/240V)

b. Rated sensed current: 30mA (120V), 10mA (220/230/240V)

Safety and Malfunction prevention device

  • Inside the chamber over heat prevention

  • Leakage breaker

  • Empty heating prevention

  • Temperature sensor breakage detection

  • Over pressure/temperature prevention

  • Memory storage for future

  • Pressure Safetyvalve

  • Low water level detection as well as water lack prevention

  • Exhaust valve knob open/closed etection

  • Insufficient sterilization detection

  • Chamber lid open/closed etection or interlock. 

Operating Environment

a. Ambient temperature range: 5 – 35 °C
b. Relative humidity: 30 – 85%
c. Atmospheric pressure: 700 – 1060hPa
d. Gradient: Within 2°